Bird-technologies 5010 Series-Manual Manual de usuario Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 27
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
11
5014 Specifications
Sensor Type Thruline two-element
line section
Elements APM or 43 series
elements. Select two
elements from the same
series, with RFL power
1/10 of FWD power.
Frequency Range
a
2 – 3600 MHz
Average Power Measurement, APM Elements, Forward
or Reflected Direction
Power Measurement
Rangea
0.1 W to 1 kW
Uncertainty
b
± 5% of reading (95% c.l.)
Peak/Average Ratio, Max 10 dB
Average Power Measurement, 43 Elements, Forward or
Reflected Direction
Power Measurement Range 0.1 W to 10 kW
Power Measurement
Uncertainty
± 5% of full scale
average power
(95% c.l.)
Peak Power Measurement, 43 Elements only, Forward
direction only
Pulse Width, Min
2 – 25 MHz
25 – 100 MHz
> 100 MHz
15 µs
1.5 µs
800 ns
Rep. Rate, Min 15 pps
Duty Cycle, Min 1 x 10
–4