Maintenance
35
Burst Average Power
Crest Factor
Complementatry Cumulative Distribution Function
(CCDF)
Physical and Environmental Specifications
RF Power Range 2 – 150 W average
Burst Width 1 µs – 50 ms
Repetition Rate, Min 15 Hz
Duty Cycle (D) 0.001 – 1
D = Burst Width / Period
Measurement Uncert. ± (6% of reading + 0.05/D W)
*
* Above 35 °C or below 15 °C add 3%
RF Power Range 0.15 – 150 W
Measurement Uncert. Linear sum of peak and average
power uncertainty
Measurement Range 0.1 – 100%
Measurement Uncert. ± 0.2%
Threshold Level Range 2 – 400 W
Level Set Accuracy As peak power uncert. + 2%
Temp, Operating –10 to +50 °C (+14 to +122 °F)
Temp, Storage –40 to +80 °C (–40 to +176 °F)
Mechanical Shock and
Vibration
MIL-PRF-28800F class 3
Humidity, Max 95% (non-condensing)
Altitude, Max 15,000 ft. (4,500 m)
Dimensions, Nominal 4.75” x 4.6” x 1.3”
(121 x 117 x 33 mm)
Weight, Max 1.2 lb. (0.55 kg)