Bird-technologies 5000-EX Manual de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 48
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
Bird Model 5000-EX Digital Power Meter
30
Bird 5011 and 5011-EF Terminating Sensors
General Corrected terminated average
power
Sensor Type Diode based terminated true
average power measurement
Frequency Range
5011
5011-EF
40 MHz to 4 GHz
40 MHz to 12 GHz
RF Power Range 10
µW to 10 mW
(–20 dBm to +10 dBm)
Max Power (Damage
Level)
2 W avg.
125 W peak for 5 µs
Peak/Average Ratio 12 dB max
Uncertainty
*
±(5% of reading ± 1 µW) (95% c.l.)
(excluding mismatch uncertainty)
(with correction factors for -EF)
RF Connector
5011
5011-EF
N Male
Precision N Male
Impedance, Nominal 50 ohms
Input VSWR:
5011
Typical
Maximum
5011-EF
Typical
Maximum
1.03 (36.6 dB return loss)
1.20 (20.8 dB return loss)
1.05 (32.0 dB return loss)
1.25 (19.1 dB return loss)
Power Supply From host instrument via cable
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 48

Comentarios a estos manuales

Sin comentarios