Bird-technologies 6151A Manual de usuario Pagina 1

Busca en linea o descarga Manual de usuario para Instrumentos de medición Bird-technologies 6151A. Bird Technologies 6151A User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
INSTRUCTION BOOK
OPERATING INSTRUCTIONS
TERMALINE
®
WATTMETER
MODEL 6151A
©
Copyright 1997 by Bird Electronic Corporation
Instruction Book Part Number 920-6151A Revision A
Termaline
®
and Thruline
®
are Registered Trademarks
of Bird Electronic Corporation
Electronic Corporation
Cleveland (Solon) Ohio US
A
Vista de pagina 0
1 2 3 4 5 6 ... 23 24

Indice de contenidos

Pagina 1 - Electronic Corporation

INSTRUCTION BOOKOPERATING INSTRUCTIONSTERMALINE®WATTMETERMODEL 6151A©Copyright 1997 by Bird Electronic CorporationInstruction Book Part Number 920-615

Pagina 2

RF Connector. ... 11Diaphragm and Coolant Oil. ... 11Sales/Repair Facility. ...

Pagina 3 - Safety Precautions

Chapter 1IntroductionPurpose andFunctionThe Bird Model 6151A Termaline RF Wattmeter is designed to measure RF powerunder non-radiating conditions. The

Pagina 4 - 6151A Caution Statements

INPUT CONNECTOR,FEMALE "N",6-11/323-15/16(100mm)(100 mm)D.C. JACK(161 mm)6-11/3213-1/64(331 mm)RETRACTABLE HANDLE3-15/16(100 mm)IN WATTSDIRE

Pagina 5

Chapter 2Theory of OperationTravelling Waves The voltage and current present on a transmission line are expressed as the sum ofthe voltages and curren

Pagina 6

dielectric coolant carries the electrically generated heat from the resistor to the wallsof the cylindrical cooling tank. The tank is encased in a set

Pagina 7 - About This Manual

Chapter 3InstallationThis chapter provides information for preparing the 6151A for use.Location Free air circulation around the wattmeter is essential

Pagina 8

To remove the meter case:1. Raise the buttons at the sides of the case, one at a time, tolift the spring above the notch in the hook. Use a flat blade

Pagina 9 - Table Of Contents

Chapter 4Operating InstructionsWARNINGExposure to RF power radiation and the possibility of RF shock or burns exists with someoperating conditions. Al

Pagina 10

MeasuringTransmitter PowerWARNINGDo not attempt to operate the Termaline Wattmeter for prolonged periods at higher thanread load levels. Damage to the

Pagina 11 - Introduction

Chapter 5MaintenanceThis chapter covers the cleaning, inspection, troubleshooting, and replacement parts.TroubleshootingFor corrections requiring repa

Pagina 13 - Theory of Operation

InspectionInspection will be necessary at six months intervals. Check for coolant oil seepagearound the radiator tank, particularly at the front and b

Pagina 14 - RF Load Section

3. Remove the diaphragm cover and lift the diaphragm fromthe back end of the radiator tank. Inspect the diaphragm.If it is no longer soft and pliable

Pagina 15 - Unlatched

Pan Asia Sales OfficeBird Electronic Ltd3A Unit 6 Tyrwhitt RoadSingapore 0820Phone: 65 299 2537Fax: 65 299 8509StorageNo special preparations for stor

Pagina 16

*Available “QC” Type ConnectorsConnector Part Number Connector Part NumberN-Female 4240-062 LT-Female 4240-018N-Male 4240-063 LT-Male 4240-012HN-Femal

Pagina 17 - Zero Adjust

Limited WarrantyAll products manufactured by Seller are warranted to be free from defects in material and workmanshipfor a period of one (1) year, unl

Pagina 18

Safety PrecautionsThe following are general safety precautions that are not necessarily related to any specific part or procedureand do not necessaril

Pagina 19 - Maintenance

WARNINGExposure to RF power radiation and the possibility of RF shock or burns exists with some operating conditions.Always be sure to turn off transm

Pagina 20

Safety StatementsUSAGEANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THEMANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION.USOEL USO

Pagina 21

ASSISTENZA TECNICALE ISTRUZIONI RELATIVE ALL’ASSISTENZA SONO PREVISTE ESCLUSI-VAMENTE PER IL PERSONALE OPPORTUNAMENTE ADDESTRATO. PER EVI-TARE PERICOL

Pagina 22

About This ManualThis instruction book covers the model 6151A Termaline Wattmeter.This instruction book is arranged so that essential information on s

Pagina 23

Bird Model 6151A Termaline Wattmetervi

Pagina 24 - Limited Warranty

Table Of ContentsSafety Precautions. .........iWarning Statements. .........iCaution Statements. .

Comentarios a estos manuales

Sin comentarios