1. Remove the power cord from the instrument if it is attached.
2. Remove the top cover from the instrument case (if supplied).
3. Unscrew and remove the filler cap from the reservoir tank.
4. Make sure the drain, located on the rear panel, is closed.
+
Note: The unit is shipped drained of coolant. The drain must be closed when fill-
ing the unit with coolant for the first time.
5. Remove the spigot from the service kit and thread it onto the coolant
container.
6. Add coolant until the level reaches the base of the filler neck. Use a
funnel to prevent coolant from spilling onto the instrument. If spilling
occurs, wipe the coolant up and allow the equipment to dry before
proceeding to step 7.
7. Connect the Calorimeter to ac main power.
8. Turn the power ON/OFF switch to on to start up the fan and coolant
pump. An alarm may sound until the coolant starts to circulate and
sufficient flow rate is achieved.
9. Run the pump for a few minutes. Refill the coolant reservoir if the
coolant level dropped.
10.Install the filler cap and wipe away any spilled coolant.
11.Reinstall the top cover onto the instrument case, if used.
9
Installation
Figure 6
Filling Coolant
Reservoir