RF Power Range* 0.1 W to 10 kW
Uncertainty
± 5% of full-scale power (95% c.l.)
Peak Power Measurement, 43 Elements, FWD only
Pulse Width, Min 2 – 25 MHz 15 s
25 – 100 MHz 1.5 s
> 100 MHz 800 ns
Rep. Rate, Min 15 pps
Duty Cycle, Min 1 x 10
–4
Uncertainty
± 8% of full-scale peak envelope
power (95% c.l.)
Match Measurement:
Measurement Range
Return Loss
Rho (
ρ)
VSWR
0 to 20 dB
0.1 to 1
1.22 to 99.99
Uncertainty Calculated from forward and
reflected uncertainty.
Twice the Avg Power Uncertainty
Settling Time, Max 2.5 seconds
Impedance, Nominal 50 ohms
Insertion Loss, Max 0.05 dB up to 1 GHz
Input VSWR, Max. 1.05:1 up to 1 GHz
Directivity, Typical* 30 dB
RF Connectors QC Type, N(F) normally supplied
Power Supply From host instrument via cable
Mechanical Shock
and Vibration
In accordance with MIL-T-28800D
Class 3
CE CE compliant. Refer to DOC for
specific standards.
Temp, Operating –10 to +50 °C (+14 to +122 °F)
Temp, Storage –40 to +75 °C (–40 to +167 °F)
Humidity, Max 95% (non-condensing)
Altitude, Max 3,000 m (10,000 ft.)
Dimensions, Nominal 2.5” x 5.0” x 2.0”
(59 x 127 x 51 mm)
Weight, Nominal 1.25 lb. (0.6 kg)
*. Exact value depends on element selected
†. Above 35 °C or below 15 °C add 2%